Flowmeter Proline 200 PROFINET with Ethernet-APL Electrical connection
Endress+Hauser 27
3.
4.
A0034171
 5 Sample graphic
Connecting cable (standard, reinforced)
3. Guide the connecting cable through the cable entry and into the connection housing (if
using a connecting cable without an M12 device plug, use the shorter stripped end of
the connecting cable).
4. Wire the connecting cable:
Terminal 1 = brown cable
Terminal 2 = white cable
Terminal 3 = yellow cable
Terminal 4 = green cable
5. Connect the cable shield via the cable strain relief.
6. Tighten the screws for the cable strain relief using a torque in the range of
1.2 to 1.7 Nm.
7. Reverse the removal procedure to reassemble the connection housing.
Connecting cable (option "mass pressure-/temperature-compensated")
3. Guide the connecting cable through the cable entry and into the connection housing (if
using a connecting cable without an M12 device plug, use the shorter stripped end of
the connecting cable).
4. Wire the connecting cable:
Terminal 1 = brown cable
Terminal 2 = white cable
Terminal 3 = green cable
Terminal 4 = red cable
Terminal 5 = black cable
Terminal 6 = yellow cable
Terminal 7 = blue cable
5. Connect the cable shield via the cable strain relief.
6. Tighten the screws for the cable strain relief using a torque in the range of
1.2 to 1.7 Nm.