Proline Promag P 100 Installation
Endress+Hauser 17
If operating outdoors:
•
Install the measuring device in a shady location.
• Avoid direct sunlight, particularly in warm climatic regions.
• Avoid direct exposure to weather conditions.
Temperature tables
SI units
T
a
[°C]
T6
[85 °C]
T5
[100 °C]
T4
[135 °C]
T3
[200 °C]
T2
[300 °C]
T1
[450 °C]
30 50 95 130 150 150 150
50 – 95 130 150 150 150
60 – 95 110 110 110 110
US units
T
a
[°F]
T6
[185 °F]
T5
[212 °F]
T4
[275 °F]
T3
[392 °F]
T2
[572 °F]
T1
[842 °F]
86 122 203 266 302 302 302
122 – 203 266 302 302 302
140 – 203 230 230 230 230
System pressure
Furthermore, install pulse dampers if reciprocating, diaphragm or peristaltic pumps are used.