Installation Proline Promass 83
8 Endress+Hauser
2.2.3 Orientation
• The direction of the arrow on the nameplate of the measuring device must match the flow
direction of the fluid.
• The following table lists the possible orientations of the measuring devices:
Vertical Horizontal Horizontal Horizontal
A0004572
A0004576
A0004580
A0015445
Transmitter at the
side
Transmitter at the
top
Transmitter at the
bottom
Transmitter at the
side
Promass A Recommended Possible (
m
) Possible (
m
,
p
) Not suitable
Promass E
Recommended
Recommended
(
n
)
Recommended
(
o
,
p
)
Not suitable
Promass F
Recommended
Recommended
(
n
)
Recommended
(
o
,
p
)
Not suitable
Promass F HT*
Compact version
Recommended Not suitable
Recommended
(
o
,
p
)
Not suitable
Promass F HT*
Remote version
Recommended Possible (
n
)
Recommended
(
o
,
p
)
Not suitable
Promass H
Recommended Recommended
Recommended
(
p
)
Recommended
Promass I
Recommended Recommended
Recommended
(
p
)
Recommended
Promass P
Recommended Recommended
Recommended
(
p
)
Recommended
Promass S
Recommended Recommended
Recommended
(
p
)
Recommended
Promass O
Recommended
Recommended
(
n
)
Recommended
(
o
,
p
)
Not suitable
Promass X
Recommended Recommended
Recommended
(
p
)
Possible
*HT = high temperature version for medium temperatures (TM) > 200 °C (392 °F)
m
Do not install the measuring device in such a way that is suspended without any support or securing unit.
n
This orientation is not suitable for fluids with entrained solids.
o
This orientation is not suitable for outgassing fluids.
p
This orientation is not suitable for low fluid temperatures.