Proline Prosonic Flow B 200 HART Technical data
Endress+Hauser 139
Electromagnetic
compatibility (EMC)
• As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
• Complies with emission limits for industry as per EN 55011
For details, refer to the Declaration of Conformity.
16.9 Process
Medium temperature range Sensor
0 to +80 °C (+32 to +176 °F)
Pressure-temperature
ratings
An overview of the pressure-temperature ratings for the process connections is
provided in the "Technical Information" document
Flow limit Select the nominal diameter by optimizing between the required flow range and
permissible pressure loss.
For an overview of the measuring range full scale values, see the "Measuring range"
section → 130
• The minimum recommended full scale value is approx. 1/20 of the maximum full scale
value.
• In most applications, 10 to 50 % of the maximum full scale value can be considered
ideal.
Pressure loss There is no pressure loss.
System pressure Sensor
Max. 10 bar (145 psi)
Thermal insulation For optimum temperature and methane fraction measurement (order characteristic for
"Sensor version", option 2 "Volume flow + Biogas analysis"), make sure that heat is neither
lost nor applied to the sensor. Thermal insulation can ensure that such heat transfer does
not take place.
Thermal insulation is particularly recommended in situations where there is a large
difference between the process temperature and the ambient temperature. This can result
in heat convection errors during temperature measurement. A further factor which can
lead to measurement errors due to heat convection is a low flow velocity.
16.10 Mechanical construction
Design, dimensions
For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section