Commissioning Proline Prosonic Flow B 200 HART
80 Endress+Hauser
Parameter overview with brief description
Parameter Prerequsite Description Selection / User
entry
Factory setting
Format display A local display is provided. Select how measured values
are shown on the display.
1 value, max. size
1 bargraph + 1
value
2 values
1 value large + 2
values
4 values
1 value, max. size
Value 1 display A local display is provided. Select the measured value that
is shown on the local display.
Volume flow
Corrected volume
flow
Corrected methane
volume flow
*
Mass flow
Energy flow
*
Methane fraction
*
Calorific value
*
Wobbe index
*
Temperature
*
Totalizer 1
Totalizer 2
Totalizer 3
Current output 1
Current output 2
*
Sound velocity
Flow velocity
Acceptance rate
*
Signal asymmetry
*
Turbulence
*
Signal strength
*
Signal to noise
ratio
*
Volume flow
0% bargraph value 1 A local display is provided. Enter 0% value for bar graph
display.
Signed floating-point
number
Country-specific:
0 m³/h
0 ft³/h
100% bargraph value 1 A local display is provided. Enter 100% value for bar
graph display.
Signed floating-point
number
Depends on country
and nominal
diameter
Value 2 display A local display is provided. Select the measured value that
is shown on the local display.
Picklist see Value 1
display parameter
None
Value 3 display A local display is provided. Select the measured value that
is shown on the local display.
Picklist see Value 1
display parameter
None
0% bargraph value 3 A selection has been made in
the Value 3 display
parameter.
Enter 0% value for bar graph
display.
Signed floating-point
number
Country-specific:
0 m³/h
0 ft³/h
100% bargraph value 3 An option was selected in the
Value 3 display parameter.
Enter 100% value for bar
graph display.
Signed floating-point
number
Depends on country
and nominal
diameter
Value 4 display A local display is provided. Select the measured value that
is shown on the local display.
Picklist see Value 1
display parameter
None
* Visibility depends on order options or device settings
10.4.8 Configuring the output conditioning
The Output conditioning wizard guides you systematically through all the parameters
that have to be set for configuring the output conditioning.