Proline Prosonic Flow B 200 HART Commissioning
Endress+Hauser 93
Parameter overview with brief description
Parameter Prerequsite Description Selection / User
entry
Factory setting
Assign simulation process variable Select a process variable for
the simulation process that is
activated.
Off
Volume flow
Corrected volume
flow
Methane fraction
Corrected methane
volume flow
Mass flow
Temperature
Energy flow
Wobbe index
Calorific value
Flow velocity
Sound velocity
Off
Process variable value One of the following options is
selected in the Assign
simulation process variable
parameter (→  93):
Volume flow
Corrected volume flow
Methane fraction
*
Corrected methane volume
flow
*
Mass flow
Temperature
*
Energy flow
*
Wobbe index
*
Calorific value
*
Flow velocity
Sound velocity
Enter the simulation value for
the selected process variable.
Depends on the
process variable
selected
0
Simulation current input 1 Switch simulation of the
current input on and off.
Off
On
Off
Value current input 1 In the Simulation current
input parameter the On option
is selected.
Enter the current value for
simulation.
3.59 to 22.5 mA 3.59 mA
Simulation current output 1 to 2 Switch the simulation of the
current output on and off.
Off
On
Off
Value current output 1 to 2 In the Simulation current
output 1 to 2 parameter, the
On option is selected.
Enter the current value for
simulation.
3.59 to 22.5 mA 3.59 mA
Frequency Output Simulation The Frequency option is
selected in the Operating
mode parameter.
Switch the simulation of the
frequency output on and off.
Off
On
Off
Frequency Value The On option is selected in
the Frequency Output
Simulation parameter.
Enter the frequency value for
the simulation.
0.0 to 1 250.0 Hz 0.0 Hz
Pulse output simulation The Pulse option is selected in
the Operating mode
parameter.
Set and switch off the pulse
output simulation.
For Fixed value option:
Pulse width parameter
(→  70) defines the
pulse width of the pulses
output.
Off
Fixed value
Down-counting
value
Off
Pulse value In the Pulse output
simulation parameter
(→  93), the Down-
counting value option is
selected.
Enter the number of pulses for
simulation.
0 to 65 535 0