Proline Prosonic Flow P 500
26 Endress+Hauser
Supported baud rate 1 200 BAUD
2 400 BAUD
4 800 BAUD
9 600 BAUD
19 200 BAUD
38 400 BAUD
57 600 BAUD
115 200 BAUD
Data transfer mode ASCII
RTU
Data access Each device parameter can be accessed via Modbus RS485.
For Modbus register information
System integration Information on system integration: Operating Instructions .
Modbus RS485 information
Function codes
Register information
Response time
Modbus data map
Power supply
Terminal assignment Transmitter: supply voltage, input/outputs
HART
Supply voltage Input/output
1
Input/output
2
Input/output
3
1 (+) 2 (–) 26 (+) 27 (–) 24 (+) 25 (–) 22 (+) 23 (–)
Device-specific terminal assignment: adhesive label in terminal cover..
Modbus RS485
Supply voltage Input/output
1
Input/output
2
Input/output
3
1 (+) 2 (–) 26 (B) 27 (A) 24 (+) 25 (–) 22 (+) 23 (–)
Device-specific terminal assignment: adhesive label in terminal cover..
Transmitter and sensor connection housing: connecting cable
The sensor and transmitter, which are mounted in separate locations, are interconnected by a
connecting cable. The cable is connected via the sensor connection housing and the transmitter
housing.
Terminal assignment and connection of the connecting cable:
Proline 500 →  27
Device plugs available
Device plugs may not be used in hazardous areas!
Device plug for connecting to the service interface:
Order code for "Accessory mounted"
option NB, adapter RJ45 M12 (service interface) →  27