Proline Prosonic Flow P 500
70 Endress+Hauser
Data logging
Manual
If the Extended HistoROM application package (order option) is enabled:
• Record up to 1 000 measured values via 1 to 4 channels
• User configurable recording interval
• Record up to 250 measured values via each of the 4 memory channels
• Export the measured value log via a variety of interfaces and operating tools e.g. FieldCare,
DeviceCare or web server
Certificates and approvals
Currently available certificates and approvals can be called up via the product configurator.
CE mark
The device meets the legal requirements of the applicable EU Directives. These are listed in the
corresponding EU Declaration of Conformity along with the standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
RCM-tick symbol
The measuring system meets the EMC requirements of the "Australian Communications and Media
Authority (ACMA)".
Ex approval
The measuring device is certified for use in hazardous areas and the relevant safety instructions are
provided in the separate "Safety Instructions" (XA) document. Reference is made to this document on
the nameplate.
The separate Ex documentation (XA) containing all the relevant explosion protection data is
available from your Endress+Hauser sales center.
Proline 500
ATEX/IECEx
Currently, the following versions for use in hazardous areas are available:
Ex db ia
Transmitter Sensor
Category Type of protection Category Type of protection
– – II2G Ex db ia IIC T6…T1 Gb
II3G Ex ec nC IIC T5…T4 Gc II2G Ex db ia IIC T6…T1 Gb
Ex ec
Transmitter Sensor
Category Type of protection Category Type of protection
– – II3G Ex ec ic IIC
II3G Ex ec nC IIC T5…T4 Gc II3G Ex ec ic IIC
Ex tb
Transmitter Sensor
Category Type of protection Category Type of protection
– – II2D Ex ia tb IIIC T** °C Db
C
CSA
US
Currently, the following versions for use in hazardous areas are available: