Proline Prosonic Flow P 500
Endress+Hauser 53
Degree of protection
Transmitter
• As standard: IP66/67, type 4X enclosure
• When housing is open: IP20, type 1 enclosure
• Display module: IP20, type 1 enclosure
Sensor
IP68, type 6P enclosure
External WLAN antenna
IP67
Vibration- and shock-
resistance
Vibration sinusoidal, in accordance with IEC 60068-2-6
• 2 to 8.4 Hz, 3.5 mm peak
• 8.4 to 2 000 Hz, 1 g peak
Vibration broad-band random, according to IEC 60068-2-64
• 10 to 200 Hz, 0.003 g
2
/Hz
• 200 to 2 000 Hz, 0.001 g
2
/Hz
• Total: 1.54 g rms
Shock half-sine, according to IEC 60068-2-27
6 ms 30 g
Rough handling shocks according to IEC 60068-2-31
Electromagnetic
compatibility (EMC)
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21) and 43 (NE43)
Details are provided in the Declaration of Conformity.
Process
Medium temperature range
Sensor version Frequency Temperature
C-030-A 0.3 MHz –40 to +100 °C (–40 to +212 °F)
C-050-A 0.5 MHz –20 to +80 °C (–4 to +176 °F)
C-100-A 1 MHz –20 to +80 °C (–4 to +176 °F)
C-200-A 2 MHz –20 to +80 °C (–4 to +176 °F)
C-500-A 5 MHz –40 to +150 °C (–40 to +302 °F)
C-100-B 1 MHz –40 to +80 °C (–40 to +176 °F)
C-200-B 2 MHz –40 to +80 °C (–40 to +176 °F)
C-100-C 1 MHz 0 to +170 °C (+32 to +338 °F)
C-200-C 2 MHz 0 to +170 °C (+32 to +338 °F)
Sound velocity range
600 to 2 100 m/s (1 969 to 6 890 ft/s)
Medium pressure range
No pressure limitation. Nevertheless, for correct measurement, the static pressure of the medium
must be higher than the vapor pressure.
Pressure loss
There is no pressure loss.