Troubleshooting
142
10.3.2 Envelope curve display in the ToF Tool - Fieldtool Package
L00-FMU90xxx-19-00-00-en-086
1. Klick on "envelope curve" (a).
2. Select the sensor (b) whose envelope curve you want to check.
3. Click on
"read curve" (c) to display a single curve
"cyclic read" (d) to display the curves cyclically.
4. Select the curves you want to check in the "Curves" window (e):
Envelope Curve
Map (= mapping of the interference echo suppression)
FAC (= Floating Average Curve)
For details refer to the operating manual of the ToF Tool - Fieldtool Package (BA224F).
(a)
(c) (d)
(b)
(e)