Commissioning
54
6.3 Preparing the basic setup
1. After the first setup the main screen appears.
However, the displayed value does not correspond
to the real level before you have performed the
basic setup. To do so, enter the main menu by
pressing "Menu" (right key).
!
Note!
In the "calibr. display" menu you can adjust the
display to your requirements (displayed values,
display format). The figure shows an example for a
2-channel instrument.
L00-FMU90xxx-07-00-00-en-049
2. Select the "level" submenu.
– Select by ↓ and ↑
– Confirm by ↵
L00-FMU90xxx-07-00-00-en-033
3. In the following submenu select the level channel
you are going to calibrate.
!
Note!
The selection "level (LVL) 2" is only available for
instruments with 2 sensor inputs or 2 current
outputs.
L00-FMU90xxx-07-00-00-en-034
4. In the following submenu select "basic setup".
This submenu contains all parameters needed for
the basic setup.
L00-FMU90xxx-07-00-00-en-035