The "Relay/Controls" menu
77
8.1.2 "relay allocation"
Use this parameter to select the relay you are going to configure.
Selection:
• All relays of the instrument version at hand
!
Note!
If a function has already been allocated to one of the relays, the name of this function is displayed
next to the relay number.
8.1.3 "relay N" (N = 1 - 6)
(Part 1: relay function)
After selecting a relay, the parameter set "relay N" (N =1 - 6) appears, which is used to configure
the relay. Initially, it contains the "function" parameter only. To configure a limit relay, proceed
according to the following steps:
1. Select the "function" parameter. The "select function" screen appears.
2. Select "limit". The "function" selection list appears.
3. Select the measured or calculated value to which the limit relay refers. The selection depends
on the instrument version and the parametrization.
!
Note!
If temperature measurement of sensor N is selected as the function, it always refers to the
temperature which has been assigned to the respective sensor in "sensor management/FDU sensor
N". Possible temperatures are:
• sensor temperature
• average of sensor temperature and temperature of an external temperature sensor
• temperature of an external temperature sensor
8.1.4 "relay N" (N = 1 - 6)
(Part 2: Limit type and switching points)
"Limit type"
Use this parameter to define the type of limit.
⇒
relay allocation RX101
relay 1
relay 2
relay 3
relay 4
⇒
relay 1 R1103
function:
⇒
relayN RX101
function: limit
limit type:
switch on point:
switch off point: