Electrical connection Prothermo NMT81
44 Endress+Hauser
bbbbbb
b
A
a
No.1
No.2
No.3
No.4
No.5
No.6
A0044611
40 Four-wire structure
A Terminal connector
a Wire from NMT81
b Wire from the temperature probe
1. Select a pair of wires (blue, red, white, and black).
2. Strip off 5 mm (0.2 in)
3. Twist both ends together and insert them into the crimp terminal [1]
4. Crimp the connected terminal with a crimp tool, a plier, or any other connection
tools.
This completes the procedure for connecting temperature signal wires.
If there are special handling requirements for the temperature probe, follow them
accordingly.
1
a
b
a
b
a
b
a
b
1.
1.
1.
2.
1.
4.
1.
3.
A0044595
41 Wire connection
a Wire from NMT81
b Wire from the temperature probe
1 Crimp terminal
NMT81 uses one common wire per connector. If your temperature probe has
additional common wires, combine them into one wire before inserting into the crimp
terminal.