Operation Prothermo NMT81
82 Endress+Hauser
Element positions
"Element positions" has two items as follows.
A0044831
84 Element positions screen
• 65 (End of probe to zero distance): Defines the distance between the physical end of the
probe and the zero level value in the tank (datum plate/tank bottom). Adjust this value
so that the absolute element positions fit to the level in the tank.
Setting range: –99 999.9900 to 99 999.9900
• 82 (Element position): Can be confirmed according to your specifications.
Setting range: –99 999.9900 to 99 999.9900
The absolute position will be changed depending on the setting of the End of probe to
zero distance. The position 1 to 24 will be automatically set based on the information
of the zero distance. However for the converter-only specification, each element
position must be manually set.
(82)
n (83)
(79)
(80)
(81)
(78)
3 (83)
2 (83)
1 (83)
(65)
0mm
A0044824
85 Element positions