Element offsets
The setting of the element offsets enables individual element 1 to 24 temperature offsets
to be activated.
Input the offset values if needed.
Setting range: –100 to 100 °C or -180 °F to 180 °F
When changing the unit of temperature (°C ↔ °F), the display and the temperature
range automatically show proper unit.
A0044832
86 Element offsets screen
• Disable: Inactivates the offset values.
• Enable: Activates the offset values.
Temperature limit
The range of the Temperature limit corresponds to the application temperature of the
order code. The range is set according to your order code when shipped.
The set values are used to calculate the bar graph on the display (option).
Setting range varies depending on the specifications.
000.00
0 10 20 30 40 50 60 70 80 90 100
1
A0044840
87 Temperature limits
1 Bar graph