RMS 621 9 Fault-finding
Endress+Hauser 53
Steam condensate temperature The measured or calculated
temperature equals the
condensate temperature of
saturated steam
Check application,
transmitters and connected
sensors.
Take steps to improve
process control - increase
temperature, decrease
pressure.
Possible inaccurate temperature
or pressure measurement; purely
a calculation of the phase change
from steam to water, that really
does not occur; inaccuracy from
set up of a temperature
compensation (approx. 1 - 3 °C)
offset.
Water: Distillation temperature The measured temperature is equal
to the distillation
temperature of water (water
evaporates!)
Check application,
transmitters and connected
sensors.
Take steps to improve
process control - decrease
temperature, increase
pressure.
Signal range violation: "Channel
ident." "Signal ident."
Output current signal is smaller
than 3.8 mA or higher than 20.5
mA.
Check if the current output is
correctly scaled.
Change the scale start and end
values.
Cable open circuit: "Channel ident."
"Signal ident.")
Input current is smaller than 3.6
mA or higher than 21 mA.
Incorrect cabling
Sensor not set to 4–20 mA range.
Functional fault at sensor
Incorrect end value set-up on
flowmeter.
Check sensor set-up
Check sensor function
Check end value of connected
flowmeter
Check cabling
Cable open circuit: "Channel ident."
"Signal ident.")
Too high resistance at the PT 100
input, e.g. caused by short circuit or
cable breakage
Incorrect cabling
PT100 sensor defective
Check cabling
Check PT100 sensor function
Min. temp. diff. undercut Overhang of the preset
temperature differential
Check actual temperature
values and the preset minimum
temperature differential.
Alarm set point violation
"Alarm set point ident." <
"Threshold" "Eng. unit"
"Alarm set point ident." >
"Threshold" "Eng. unit"
"Alarm set point ident." >
"Gradient" "Eng. unit"
"Alarm set point ident. <
"Gradient" "Eng. unit"
"User defined message"
Alarm set point violation (higher or
lower than).
( Setup, Alarm set points, Chap.
6.4.3)
Acknowledge alarm message if
the function “Alarm set
point/Event text/Display and
acknowledge“ has been set up.
( Setup, Alarm set points,
Chap. 6.4.3)
If required check application
If required reset alarm set points
Display message Cause Cure