Operating the Analyzer
Operator’s Manual 2
–9
b. If an automatic validation or Mode 7 has been processed:
• Date is the time of the last validation.
• 1: is for Validation 1.
• P or F equals ‘Pass’ or ‘Fail’ for the validation result.
• 1000.00ppmv is the concentration of the last
validation result in the user’s selected engineering
units. If the value is from a Mode 7 validation, then
it is the average validation value for the time period
during which the Mode 7 was run.
• Rng:1000.0 to 1000.0 is the minimum and
maximum concentration value, in the user’s
selected engineering units, during the last
validation time period.
For systems set up for dual validation, the following screens may be displayed:
a. If an automatic validation or Mode 7 or Mode 8 has not yet
been processed:
b. If an automatic validation, Mode 7 or Mode 8 has been pro-
cessed:
• Date is the time of the last validation.
<VALIDATION RESULTS>
Date: 17-04-10 13:00
1:P 1000.00ppmv
Rng:1000.0 to 1000.0
<VALIDATION RESULTS>
Date: 17-04-10 13:00
1: NO DATA
2: NO DATA
<VALIDATION RESULTS>
Date: 17-04-10 13:00
1:P 1000.00ppmv
2:P 1000.00ppmv