NS 5.14 Firmware
214 4900002262 rev. B 3-23-21
Table 2-1 Typical values for parameter setpoints (Continued)
Parameter Setting Function
Validation Allowance 0.0 to 100.0
Default: 10
Sets acceptable deviation
(in %) for validation.
Zero Val Tolerance 0.0 to 1000000
Default = 0
Sets maximum acceptable zero
measurement reading during
validation routine.
Daily Validation 0 or 1
Default: 0
Turns daily autovalidation on or
off.
Val Interval 1 to 400
Default: 1
Interval (in days) between
validation cycles.
Val Start Time 0 to 23
Default: 9
Sets the hour of the day for
validation.
Start Validation 0 or 1
Default: 0
Initiates validation cycle.
4-20 mA Val Action 0 or 1
Default: 0
Sets the current loop mode
during validation.
ValPerm Constant Kp 0 to 1000000
Default = 0
The system constant for the
perm tube.
ValPerm Rate Rp 0 to 1000000
Default = 0
The calibrated permeation rate
for the perm tube.
Cancel Val Alarms 0 or 1
Default: 0
Resets the validation alarms
and relays.
Val Auto DumpSpectrm 0 or 1
Default: 0
Sets the analyzer to dump
spectrum information during a
validation measurement.
Operator Password 0 to 9999
Default: 0
Password for Operator
Parameter section.
Operator Parameter01
to Operator
Parameter20
Parameter Index
Default: 0
Parameter setup for Operator
Parameter section.