Maintenance  13. Calibration 
 
LS20-B  Rev.4  173 
)-3  Move the Manipulator by hand to a rough 
position/posture of the calibration 
point data. 
)-4  Create the data from the calibration point data. 
Enter and execute the following command in [Command Window]. 
(In this example, P1 is used as the calibration point data.) 
> Calpls Ppls(P1,1), Ppls(P1,2), Ppls(P1
Ppls(P1,4) 
-5  Move the joint to the specified point using a motion command. 
For example, when the specified point data is “P1”,   
execute “Jump P1:Z(0)” from [Jog & Teach]. 
*  The joint NOT being calibrated moves to the original position.   
-6  Accurately  align  the  joint* being calibrated  to the specified point using 
commands. 
* You must move Joint #3 and #4 to the position when calibrating Joint #4. 
Select the jog mode [Joint] from [Jog & Teach] to execute the jog motion. 
-7  Execute the calibration. 
Enter and execute one of the following commands in [Command Window] 
according to the joint being calibrated. 
Joint #1: 
>Calib 1 
Joint #2: 
>Calib 2 
Joint #3: 
>Calib 3 
Joint #4: 
>Calib 3, 4 
 
 
-1  Move the Manipulator to another point to confirm that 
position.     
If it does not move to the same position, re-calibrate using another point.  
must set the point again 
if reproducibility cannot be assured through calibration.