11 / 56
3 • Unit description
80116-261
2018-07
CHAPTER 3
Unit description
Unit variants
The EIP 2 (Erbe Irrigation Pump) is available in the following unit variants:
• 230 V with ECB socket
1
• 230 V without ECB socket
• 120 V without ECB socket
• 100 V with ECB socket
• 100 V without ECB socket
Factory setting
The EIP 2 has the following settings when it is delivered by the manufac-
turer:
• Quick activation button I = 30%
2
• Quick activation button II = 50%
• Quick activation button III = 80%
• Automatic pressure control activated
for units up to software version 1.0.0
• Automatic pressure control deactivated
(for units from software version 1.1.0)
1. ECB = Erbe Communication Bus for connecting the EIP 2 to a VIO electrosurgical unit.
2. Flow in % relative to the maximum flow of 500 ml/min.
IMPORTANT! Once settings have been changed and saved they still apply even after
the unit has been restarted. A reset (= switching the unit back to the
factory setting) is not possible.