9 • Maintenance and servicing
80116-261
2018-07
User manual and visual inspections
• Test specimen and accessories (where enclosed) undamaged exter-
nally.
• User manual present.
• All labels on the test specimen (conformity declaration mark, rating
plate, and all wording) present and readily legible.
Tests to be conducted in accordance with the national
specifications and regulations
Grounded conductor test • Ground terminal to chassis.
• Ground terminal to potential equalization pin.
Leakage current
measurement
• Ground leakage current, normal condition (N.C.).
• Ground leakage current, single-fault condition (S.F.C.).
• Patient leakage current, normal condition (N.C.).
• Patient leakage current, single-fault condition (S.F.C.).
Performance tests
Testing tools
Test setup
• The test specimen is connected to the power supply via the power
cord.
Test procedure
Power switch 1. Check power switch for smooth operation. The power switch must be
easy to operate and must neither stick nor scrape.
2. Press power switch. The power switch must snap into the "ON" posi-
tion and the test specimen must perform a system start.
Control buttons 1. Check all the buttons on the control panel to make sure they are
operating properly. Press each button at least twice.