7 • Service programs
28 / 56
80116-261
2018-07
Service program S3 Continuous running / Burn-In
Description
Service program S3 checks whether EIP 2 can withstand a substantial con-
tinuous load.
If the service program ends successfully, the message "CS3" (Complete
Service Program
3) is output on the display. If a fault occurs during execu-
tion, the message "ES3" (Error Service Program 3) appears.
Both messages can be canceled by pressing the activation button.
If a fault occurs during execution, its cause can be localized in operating
mode. To do this, abort the service program. The unit then automatically
switches to operating mode and the display shows an error message.
Run time
10 minutes.
Set value
Fixed at 80 %.
To activate the service program
Press the activation button.
To abort the service program
Press quick activation button III.