10 • Maintenance and servicing
112 / 144
Art.-Nr.: 80116-271
04.06
HF power output COAGULATE
Testing and measuring equipment
BIPOLAR FORCED COAG Test set-up
Fig. 10-11
• The test specimen is connected to the power supply via the power
cord.
• The levels of power are determined with the HF power meter. The
measuring cables are plugged into the HF power meter direct.
• The dual-pedal footswitch with ReMode is connected.
ERBE
Art. No.
Description
20192-127
20192-110
Patient cable AE
or
Patient cable AE, international
20190-045 Electrode handle ICC/ACC
20194-070
20194-075
Patient cable NE
or
Patient cable NE, international
20100-033 Adapter cable NE
20196-045
20196-053
Bipolar cable
or
Bipolar cable, international
20100-034 Adapter cable bipolar
– HF power meter
20189-101 Dual-pedal footswitch with ReMode
ERBE
HF-Leistungsmessgerät
Bipolar