9 • Troubleshooting
92 / 144
Art.-Nr.: 80116-271
04.06
9-Errors
Recognizing module: 9 = ERBE Irrigation Pump EIP 2
Recog-
nizing
module
Error
code
Additional
information
Description Remedy
9 80 CRC Error. CRC error:
Occurs if the EIP 2 CRC
monitoring procedure per
-
formed at unit startup dis-
covers an error (e.g.
turnover of a bit in the flash
memory or error during
software download).
Reprogram the EIP 2 main
board or replace the main
board.
9 82 Pump lid open at
ON.
User error:
The pump lid of the EIP 2 is
not closed at the start of
activation or is opened dur
-
ing activation.
Close the pump lid before
activation or do not open it
during activation.
Unit error:
The Hall sensor circuit
board is not connected to
the main board or the Hall
sensor circuit board is
faulty.
Check jumper J35 between
the Hall sensor circuit
board and the main board
or replace the Hall sensor
circuit board.
9 84 Time Out Error. User error:
Activation of the EIP 2 was
maintained for more than
20
seconds. Maximum acti-
vation time for the EIP 2 is
20
seconds.
Discontinue activation.
Then the EIP
2 can be
reactivated.
9 88 UeWS Error. The motor control monitor-
ing circuit is faulty.
Replace EIP 2 main board.
9 90 No current or no
load.
During activation there is
no measurable current
input to the motor.
Check jumper J10 between
EIP
2 main board and
motor or replace the EIP
2
main board and/or motor.
9 A0 Current too high at
ON.
The current input to the
motor is too high. The
motor or pump head is
jammed or faulty.
Check the motor and pump
head. If necessary, replace
the motor and/or pump
head.
9 C0 Motor control faulty
(voltage too high at
ON).
The measured voltage is
too high. The motor control
is faulty.
Replace EIP 2 main board.