9 • Troubleshooting
Art.-Nr.: 80116-271
04.06
D-Errors
Recognizing module: D = Smoke evacuation system IES 2
Recog-
nizing
module
Error
code
Additional infor-
mation
Description Remedy
D 80 ECB fault. Check IES 2 and VIO elec-
trosurgical unit software
versions for compatibility.
D 81 Excess temperature
(>
60 °C).
Switch off IES 2 and wait
until the temperature in the
interior of the unit has
returned to normal.
D 82 Insufficient temperature
(<
0 °C).
Switch off IES 2 and wait
until the temperature in the
interior of the unit has
returned to normal.
D 83 Software error
(check
total).
Reinstall software.
IMPORTANT! Software
may only be re-installed
and updated by ERBE Ele
-
ktromedizin Tübingen.
Send unit for repair.
D 86 Pressure sensor error Restart IES 2. Replace
control board.