SpiroSphere
®
ECG Instructions for Use
Page 20/136
Version 02.02 • 04AUG2022
Calibration Check
The sensor check consists of a calibration check as well as a linearity check.
Following screen appears:
Ensure a new ERT PT (with mouthpiece removed) is connected to the 3 L calibration syringe via
an adapter (as shown).
In order to perform a calibration check, tap on <Calibration Check>.
Tapping on the "i"- symbol
will display information on the
respectively selected check
type.
Tap <Sensor Check> to perform a sensor check.
Sensor Check
The ERT PTs included with the delivery are pre-calibrated as part of manufacture.
A sensor check can be performed to conrm accurate measurement data.
An automatic zero adjustment is performed.