EasyManua.ls Logo

ERT SpiroSphere - Linearity Check

ERT SpiroSphere
136 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SpiroSphere
®
ECG Instructions for Use
Page 22/136
Version 02.02 • 04AUG2022
End the linearity check by tapping on <Save>.
Screen display after a total of 9 syringe strokes:
better than ± 3.5 %
X
worse than ± 3.5 %
In order to perform a linearity check, tap on <Linearity Check>.
Proceed as described in the "Calibration Check" section.
During a linearity check, volume accuracy at different ows is tested. Three syringe strokes at a
low, three at a mid-range ow and three at a high ow are required.
With each syringe stroke, the volume accuracy should be within ± 3.5 %.
Linearity Check
It is important to pump without interruption from impact to impact. The rst pump stroke
is not relevant and will be discarded. Three pump strokes are required in each of the
following ow levels; low, mid and high range.
1 syringe stroke = pump twice, i.e. from impact to impact.

Table of Contents

Related product manuals