3.1.3UsingSashWindow
Thesashwindowshouldbefullyclosedwhenthehoodisnotinuse.Thishelpskeeptheworkzone
interiorclean.
Thesashwindowshouldalwaysbeinthenormaloperatingheightatalltimeswhenthehoodisin
use.Evenifthecabinetisleftunattended,
buttheblowerison,thesashwindowshouldneverbe
movedfromthenormaloperatingheight,unlessduringloadingorunloadingofmaterials/apparatus
intothehood.
Thealarmwillbeactivatedwheneverthesashwindowismovedfromthenormaloperatingheight.
Thesashwindowmaybe
openedtoitsmaximumpositionforthepurposeofloading/unloadingof
materials/apparatusintothehood.Whenthesashwindowisfullyopened,thealarmsoundmaybe
mutedbypressingMUTEbutton.Lightcanbeturnedontofacilitatecleaning.
3.2StartingandShuttingDowntheDuctlessFumeHood
3.2.1TurningontheHood
1. Raisethesashtotheindicatednormaloperationalheight(READYstate).
2. Turnonthelampbypressingthelampbutton.
3. TurnonthefanbypressingtheFANbutton.InputtheFanPINifasked(default:0001).Thiswillstart
thewarmupprocedure(default:3minutes).All
buttonsaredisabledduringwarmupperiod.
4. Thehoodisreadyforwork.
3.2.2TurningofftheHood
1. TurnoffthefanbypressingtheFANbutton.InputtheFanPINifasked(default:0001).Thiswillstart
thepostpurgeprocedure(default:0minute).Allbuttonsaredisabledduringpostpurgeperiod.
2. Lowerthesashtothefullyclosedposition.Thesashcanbeloweredimmediately
afterturningoffthe
fanasitwillnotinterruptthepostpurgeprocedure.
3.3OperatingtheDuctlessFumeHood
3.3.1WorkingintheHood
Checkthelabelontheductlessfumehoodtoseewhatchemicalsthehoodisintendedforandonly
usethehoodforanyprocedureinvolvingsuchchemicals.
Ensuretheexhaustisoperatingbeforecommencingwork.
Afteralltheapparatuses/itemshavebeenarranged,allowtheblowerto
runforanother3minutesin
ordertopurgeworkzoneofcontaminants.
Minimizeroomactivitysincetheseexternalairflowdisturbancesmayadverselyaffectthehood’s
internalairflow,impairingthecontainmentcapabilitiesofthefumehood.
Keepyourheadoutsideofthehood.
Workasfarinto
thehoodaspossibleandwithslow,deliberatemovements,tominimizeairflow
disturbances.
Workwiththesashasfullyloweredaspossible,utilizingthesashasanaturalbarrier.
Donotusethishoodasastoragearea.Itemscanblockairflowandinterferewithcontainment.
Ifperformance
issuspected,oranairflowalarmistriggered(ifinstalled),terminateusage,closethe
sashcompletely,andceasework.
Donotletorganicchemicalsevaporateinthehooduseaproperwastebottle.Donotleave
uncappedbottlesofchemicalsorwasteinahood.
Certifythis
hoodannuallytoverifyairflowvelocity,smokepatternsandcontainment.
Performroutinemaintenanceinaccordancewiththemanufacturer’sinstructions.