EasyManua.ls Logo

Esco Ascent max - Installation Procedures

Esco Ascent max
71 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
DuctlessFumeHoods
4.1.4ExhaustRequirements
Theexhaustfilterareaisespeciallysusceptibletodisruptiveaircurrentsorairdrafts.Aclearanceofatleast30
cm(1’)isrecommendedbetweenthehighestpointofthecabinetandtheceiling.Ifthedistanceislessthan30
cm(1’),theairflowalarmsystemmayneedre‐calibration.
Infact,forproperexhaustfilterleakscanning
purposes,aminimumclearanceof50cm(1’8”)isrecommended.
4.1.5ElectricalRequirements
• Thepowerratingforeachmodelisshowninsection3.2TechnicalSpecificationSummaryTablein
ProductSpecificationSection.Ensurethattheoutletisratedaccordingly.Thehoodwillnotwork
properlyormayevenbedamagedifitispoweredbyanincorrectsource.
• Thehood’smaximumvoltage
fluctuationis±2%ofnominalvoltage,otherwiseinstallpowerstabilizer.
• Thehoodshouldbeconnectedtoanunobstructeddedicatedpoweroutlet(s).
• SurgeprotectionandUPSarestronglyrecommendedforbetterprotection.
• Allelectricalwiringsforfluorescentlamp,blowerandtheSentinel™controllerareinternallywired
andconnected
toasinglepointjunctionboxforhookupbyaqualifiedelectrician.
• Thepowercableislocatedontherighthandsideofthefumehoodandthecordis2.5mlong.When
preparingtheinstallationsite,trytoensuretheoutletislocatedtotherightofthe
fumehoodfor
easeofaccess.
4.1.6ServiceLineRequirements
• Allservicelinesshouldbeinstalledbyasuitablyqualifiedandcertifiedengineer,inaccordancewith
allapplicablelocal,stateandgovernmentregulations.
• Servicelineattachmentsshouldbeequippedwithanemergencyshutoffvalvethatcanbeaccessed
quicklyandwithease,shouldtheneedarises.
• You
shouldcheckwithyourlocalserviceinstallerastowhetherthereisaneedtoinstallpressure
regulatorstoreducethelinepressure.
• Yourductlessfumehoodcanaccommodateservicefixturesontheleftorrighthandsideofthe
cabinet.Makeallowanceforthepositioningofservicelines
whenplanningtheinstallationsiteto
ensureeaseofaccesstoemergencyshutoffvalves.
4.2Installation
4.2.1ConnectingtheElectricalSupply
• Beforeconnectinganyelectricalwiringtothefumehoodstructure,firstrefertotheseriallabelfor
theproperelectricalcharacteristics.
• Review the electrical wiring diagrams and instructions in Appendix section prior to installation. All
wiringshouldbedoneinaccordancewiththeNationalElectricalCode.
• Connectthe
suppliedelectricalcordtotheinputonthetopofthefumehood.Makesurethecable
connectorisseatedfirmlyinthesocket.
4.2.2SafetyLabelsontheFumeHood
• Anyoneusingthefumehoodshouldfamiliarizethemselveswiththevariouslabelsdisplayedinandon
thefumehood.
• Itisveryimportantthatusersarefamiliarwiththemeaningsofproductlabelingbeforeattemptingto
usetheunit.
4.2.3PreliminaryCleaning
Wipetheinteriorandexteriorofthefumehoodwithwateroramildhouseholddetergent.

Table of Contents

Related product manuals