18
Thedistancefromtheaperturetotheapertureofan
opposingfumehood,fumecupboard,etc.shouldbe
inexcessof3m(10’)
Youshouldnotpositionthefumehoodinalocation
wherethereisanopposingwall(orotherobstruction
likelytoaffectairflow)within2m(7’)ofthefront
aperture.
Aswithwalls,anylargeobstructionsuchasapillarorcolumnprojectingbeyondtheplaneofthefront
apertureshouldnotbewithin30cm(1’)ofthesidesofthefumehood.
4.1.1.3RelocatingtheHood
Normallyductlessfumehoodarerarelymovedoncetheyareintheirideal
positions,butiftheneedariseto
relocateorevenrepackagetheunit,thenhereissomeconsiderations:
• Beforemovingthefumehood,remembertosecureallmovingparts,ex:sash.
• Fumehoodisheavy,becarefulwiththeweightdistribution.Usuallyitwilltake6ormore
peopleto
moveit.
Forrepackaging:
• Boltthefumehoodtothepallet.
• Strapthefumehoodbodydowntothepallet.
• Repackageasnecessary.Ifpossible,useoriginalpackaging.
• Whenmovingthefumehood,usematerialhandlingequipmentandliftthepallet.
4.1.2EnvironmentalRequirements
• Indooruseonly
• Altitudeofupto2000meter(6600’)
• Relativehumiditybetween20%–90%
• Temperaturebetween18
o
C–30
o
C(65
o
F–86
o
F)
4.1.3SupportRequirements
• Thesupport/cabinetryshouldbeleveled
• Thesupport/cabinetrymustbeabletowithstandtheweightofthecabinetandanyapparatuswithin
• Thesupport/cabinetrymustbeergonomics
• EscorecommendsEscosupportstandwithlevelingfeetforBSC
• WithoutEsco’ssupportstand,Escocannotguaranteethecabinet’sresistanceagainsttipping
andhencetheuserwouldbesolelyresponsibleforensuringthatthecabinetissecurely
fastenedtothirdpartysupport/cabinetry.
• Theuseofnon‐levelingfeetEsco
supportstandwillnullifythethirdpartycertification(NSFor
TÜV)thatthecabinetmayhave,becauseonlyEscolevelingfeetsupportstandwasused
duringcertification.ThemaximumNSFapprovedleglevelersadjustmentis50mm(2”).