DuctlessFumeHoods
Chapter6–Recertification&Calibration
Ductlessfumehoodgenerallyrequirere‐certification,when:
• Thehoodisre‐located
• Hood’sperformanceissuspected
• Afterfilterorblowerreplacement
• Atleastonceayear
Duringrecertification:
• Cabinetairflowvelocitiesandflowpatternsareverifiedagainstthemanufacturer’sspecificationsand
relevantinternational
standards
• Filtersarescan‐testedtoensuretheydonotleak
• Operatorcomforttestsmaybeperformed
• Ifairflowvelocitiesarefoundtobeoffsetpoint,adjustmentsaremadeaspartofthecertification
processbeforefinalvaluesarerecorded.
Airflowalarmcalibration,when:
• The
causeoftheairflowalarmcannotbedetermined
• Re‐certificationindicatesthedisplayedairflowdeviatesby>0.02m/s(4fpm)fromactualmeasured
velocities.
6.1CertificationFlowchart