5
1
3
2
Follow the same procedure (5)~(7) of the continuous keratometry and refractometry [K/R] mode.
Example of printout:
*: Confidence of result
Level Quality
A
B
C
D
E
(Hight)
(Low)
3. Keratometry - [KER] mode
Only the radius of curvature of the cornea can be measured in [KER] mode.
Don't measure the base curve of contact lens in this mode. Measurement error will occur when it is
measured in [KER] mode.
Enter [KER] mode.
Press [Mode] button until [KER] is displayed on the right side of the display.
Follow the same procedure (2) and (3) of the continuous keratometry and refractometry [K/R] mode.
Measurement.
Press the measuring button.
Measurement will be continuously performed when the measuring button is kept pressed.
The measured results will be displayed on the monitor.
In case of the continuous measurement, previous measured result will be displayed.
When measured results are displayed, changing at PAGE 1 of [Setup] mode renders each cornea
measured result of R1/R2/AX > K1/K2/AX > AR/CY/AX displayed in turn.
USER MANUAL > MEASUREMENT
AKR300 - Auto Kerato Refractometer > V3 - 04-2020
28

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Essilor Instruments AKR 300 and is the answer not in the manual?

Essilor Instruments AKR 300 Specifications

General IconGeneral
BrandEssilor Instruments
ModelAKR 300
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals