54
55
3
32
5
4
Pump is being prepared
The pump will switch on for a few
seconds while the reservoir is prepa
red for drug delivery. At the end of this
process, the liquid should be visible at
the opening of the infusion set.
If the pump does not remain
upright during preparation, air
could remain in the reservoir,
leading to incorrect dosing of the
drug.
Stand pump with the infusion set in
docking station
The pump must remain upright during
the next step to ensure that any air is
removed from the reservoir. To do so,
place your pump with the connected
infusion set in the docking station.
Then press
.
Observing and stopping the priming
cycle
Observe the infusion set during the
priming cycle. It takes a few seconds
until liquid becomes visible in the
tube. The drug is then slowly pushed
in the direction of the needle on the
infusion set.
Pump ready
Pump and reservoir are now ready for
use. If you wish to prime the infusion
set, then press
Yes .
Press
No to skip this step. The
infusion set will not be primed.
Proceed with Step 8 "Put on infusion
set".
It may be necessary to perform
multiple priming cycles until the
liquid is visible in the tube.
If no liquid is visible after multiple
priming cycles, check to ensure
that the reservoir has been filled
properly through the viewing
window. If necessary, repeat the
filling process with a new reservoir
and return to Step 1.