EasyManua.ls Logo

Ever Pharma D-mine Pump - Daily Routine

Default Icon
83 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
58
59
4
3.4
STARTING AND STOPPING
DRUG DELIVERY
Stopping drug delivery
Press and hold
Stop and do not
release the button until you hear an
acoustic signal (approx. 3seconds).
Your pump will stop automatically
if the reservoir is empty or an error
occurs. Following an alarm, you
always have to restart the delivery
of apomorphine.
You can start and stop delivery of the
drug using the function buttons. An
interruption is only necessary if you
wish to
change the reservoir
change the infusion set
change the rechargeable battery
change the time
change delivery settings
take o the device
Starting drug delivery
Press and hold
Start and do and do
not release the button until you hear an
acoustic signal (approx. 3seconds).
Check the viewing window from
time to time to verify whether
the filling level of the reservoir is
decreasing. In this way, you can
tell whether the delivery function
is
enabled and functioning properly.
4.1 Menu control
4.2 Delivering a bolus
4.3 Changing the infusion set
4.4 Changing the reservoir
4.5 Viewing the delivery settings
4.6 Viewing the history
4.7 Changing the rechargeable battery
4.8 Taking o and recharging the pump
DAILY ROUTINE

Table of Contents