7
Warnings on the pump require the user to respond within a defined time period.
In the event of a warning, drug delivery continues unchanged. The pump will emit
an acoustic signal to notify the user of the message. The warning will remain
displayed until you acknowledge the message with
.
7.3
WARNINGS
7.4NOTICES
Notices are messages concerning
the condition of the pump. They have
no impact on the progression of the
delivery used earlier. Therefore, the
device will only display them on the
screen and will not emit any acoustic
signals.
The tables on the pages below descri
be the various notices and trouble
shooting measures.
Notices appear only once. If you
do not acknowledge a notice, the
notice is only displayed until the
screen is switched o.
Warning Reason for alarm What should I do?
Battery low
The "Battery low" warning will appear
when the remaining battery capacity falls
below 20%.
Press
to acknowledge the warning
and replace the rechargeable battery at the
next opportunity in accordance with the
instructions in Section
4.7
.
Reservoir
low
The "Reservoir low" warning will appear
when the amount of drug remaining in the
reservoir reaches the set warning time.
Press
to acknowledge the warning and
change the reservoir as soon as possible
in accordance with the instructions in
Section
4.4
.