SOURCES OF INTERFERENCE
The EVER Pharma Dmine
®
Pump has been tested as a class B, group 1 medical device in accordance with
IEC606011‑2: 2014. It is intended for use in clinics, hospitals and domestic environments.
The EVER PharmaDm ine
®
Pump delivers apomorphine. Using the device in the vicinity of strong electromagnetic
fields can impair or impede performance. In such cases, the EVER PharmaDmine
®
Pump may signal that an error
has occurred.
Avoid using the EVER PharmaDm ine
®
Pump in the vicinity of active electrosurgical instruments and within RF
shielded rooms for MRT at medical facilities where the electromagnetic wave intensity is high.
The use of the EVER PharmaDmine
®
Pump in the vicinity of or in connection with other devices should be avoided
because this can lead to malfunctions.
The use of accessories, converters or cables that are not intended for or supplied with the EVER PharmaD‑mine
®
Pump can result in increased electromagnetic emissions, reduced electromagnetic immunity, and malfunctioning
of the pump.
ELECTROMAGNETIC EMISSIONS
Portable RF communications equipment (including peripheral devices, such as antenna cables and external antennae)
should not be used at a distance of less than 30cm or 12inches from any part of the EVER PharmaDmine
®
Pump,
including cables. Such equipment includes mobile phones, wireless phones, and wireless computer devices.
Otherwise the performance of the EVER PharmaD‑mine
®
Pump may be reduced.
RF emissions CISPR 11
Group 1 Class B
Harmonic emissions
IEC 61000‑3‑2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000‑3‑3
dmax < 4% dmax < 4%