Loading...
152
153
10
ELECTROMAGNETIC IMMUNITY
IEC 60601test level Compliance level
Electrostatic discharge (ESD)
IEC 61000‑4‑2
+/‑ 8kV contact
+/‑2kV, +/‑4kV, +/‑8kV +/‑15kV air
+/‑ 8kV contact
+/‑2kV, +/‑4kV, +/‑8kV +/‑15kV air
High‑frequency electromagnetic
fields
IEC 61000‑4‑3
10 V/m
80 MHz ‑2.7 GHz
80% AM at 1kHz
10 V/m
80 MHz ‑2.7 GHz
80% AM at 1kHz
27 V/m
385 MHz PM 18Hz
27 V/m
385 MHz PM 18Hz
28 V/m
450 MHz PM 18Hz
28 V/m
450 MHz PM 18Hz
9 V/m
710 MHz PM 217 Hz
745 MHz PM 217 Hz
780 MHz PM 217 Hz
9 V/m
710 MHz PM 217 Hz
745 MHz PM 217 Hz
780 MHz PM 217 Hz
28 V/m
810 MHz PM 18 Hz
870 MHz PM 18 Hz
930 MHz PM 18 Hz
28 V/m
810 MHz PM 18 Hz
870 MHz PM 18 Hz
930 MHz PM 18 Hz
28 V/m
1720 MHz PM 217 Hz
1845 MHz PM 217 Hz
1970 MHz PM 217 Hz
28 V/m
1720 MHz PM 217 Hz
1845 MHz PM 217 Hz
1970 MHz PM 217 Hz
IEC 60601test level Compliance level
Recommended separa
tion distance
RF interference
currents
IEC 61000‑4‑6
3 Vrms
150kHz ‑ 80 MHz
3 Vrms
150kHz ‑ 80 MHz
𝑑 = 1.2 𝑃
80 MHz - 800 MHz
Radiated RF
disturbances
IEC 61000‑4‑3
10 V/m 80 MHz ‑ 2.7 GHz 10 V/m 80 MHz ‑ 2.7 GHz
𝑑 = 1.2 𝑃
80 MHz - 800 MHz
𝑑 = 2.3 𝑃
800 MHz - 2.5 GHz

Table of Contents