High‑frequency electromagnetic
fields
IEC 61000‑4‑3
28 V/m
2450 MHz PM 18Hz
28 V/m
2450 MHz PM 18Hz
9 V/m
5240 MHz PM 217 Hz
5500 MHz PM 217 Hz
5785 MHz PM 217 Hz
9 V/m
5240 MHz PM 217 Hz
5500 MHz PM 217 Hz
5785 MHz PM 217 Hz
Fast transient electrical
disturbances/bursts
IEC 61000‑4‑4
+/‑ 2kV
100kHz repetition rate
+/‑ 2kV
100kHz repetition rate
Signal input/output
+/‑1kV
100kHz repetition rate
Surge
IEC 61000‑4‑5
+/‑1kV line‑to‑line
+/‑2kV line‑to‑ground
+/‑0.5kV line
+/‑1kV line‑to‑line
+/‑2kV line‑to‑ground
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000‑4‑11
Voltage dips:
0% UT; 0.5cycle
at 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
0% UT; 1cycle at 0°
and
70% UT; 25cycles (50 Hz) / 30cycles (60Hz) at 0°
Power failure 0% UT 250cycles (50Hz),
300cycles (60Hz)
Voltage dips:
0% UT; 0.5cycle
at 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
0% UT; 1cycle at 0°
and
70% UT; 25cycles (50 Hz) / 30cycles (60Hz)
at 0°
Power failure 0% UT 250cycles (50Hz),
300cycles (60Hz)
Power frequency magnetic fields
(50/60 Hz)
IEC 61000‑4‑8
30 A/m 30 A/m
Recommended separation distance between portable and mobile RF
communications equipment and the Ever Pharma D‑
mine
®
Pump
Separation distance according to frequency of transmitter [m]
Rated maximum output
of transmitter [W]
150kHz to 80 MHz
outside the ISM bands
80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01
0.12 0.12 0.23
0.1
0.38 0.38 0.73
1
1.2 1.2 2.3
10
3.8 3.8 7.3
100
12 12 23
The EVER PharmaD‑mine
®
Pump is suitable for use in an electromagnetic environment in which radiated RF disturbances
are controlled. Customers and users of the EVER PharmaD‑mine
®
Pump can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
EVER PharmaD‑mine
®
Pump according to the maximum output power of the communications equipment.