INSERTING THE RESERVOIR
To use the reservoir, you will need:
a new vial with apomorphine from EVER Pharma
a new reservoir
a new infusion set
a docking station for maintaining the pump upright
Check whether you have received the correct drug, a new
reservoir and infusion set, and whether the vial is new.
Only use completely filled apomorphine vials in accordance
with your prescription. The use of expired, incompletely
filled or damaged vials can impair the ecacy of the
therapy
and endanger your health.
Never use a reservoir if its packaging is damaged. If the
packaging is damaged, the reservoir is no longer sterile and
could be contaminated.
Never use a reservoir more than once.
The rechargeable battery does not have to be fully
charged
in order to use the reservoir.