Eversense XL CGM User Guide
18
Clarke Error Grid Accuracy by Duration of Wear
Percent values are reported to single precision.
Overall, 84.3% of CGM readings fell within zone A. This indicates CGM readings were in close agreement with
reference values for the great majority of readings. Also, there were no real dierences in zone A percentages as the
duration of wear increased.
Period of Wear
Number of paired
CGM-YSI Readings
Clarke Score
A B C D E
Days 1 - 30 10761 84.4% 14.7% 0.1% 0.9% 0.0%
Days 1 - 60 15143 84.8% 14.4% 0.0% 0.8% 0.0%
Days 1 - 90 16572 84.8% 14.5% 0.0% 0.7% 0.0%
Days 1 - 120 19244 84.5% 14.7% 0.0% 0.8% 0.0%
Days 1 - 150 20219 84.5% 14.7% 0.0% 0.8% 0.0%
Days 1 - 180 21527 84.3% 14.9% 0.0% 0.7% 0.0%