18
Consensus Error Grid Accuracy by Glucose Level
Overall, 86.4% of CGM readings fell within zone A. This indicates CGM readings were in close agreement with
reference values for the great majority of readings.
Zone Frequency Percent
A 18598 86.4%
B 2853 13.3%
C 75 0.3%
D 1 0.0%
E 0 0.0%
Total 21527 100.0%