81
Eversense XL CGM User Guide
9
App DisplayDescription
Out of Range Low Glucose*
Appears when your glucose reading is lower than
2.2 mmol/L.
No glucose reading can be displayed (only LO is
displayed on the MY GLUCOSE screen).
Out of Range High Glucose*
Appears when your glucose value is higher than
22.2 mmol/L.
No glucose reading can be displayed (only HI is
displayed on the MY GLUCOSE screen).
Actions
Confirm your glucose value with a blood glucose
meter test before making a treatment decision.
Once the sensor glucose value is at or higher than
2.2 mmol/L, glucose readings will resume on the
display.
Measure your glucose manually by using your blood
glucose meter. Always confirm your glucose value
with a blood glucose meter test before making a
treatment decision.
Once the sensor glucose value is at or lower than
22.2 mmol/L, glucose readings will resume on the
display.
Alert Descriptions and Actions (continued)
Alerts