84
Eversense XL CGM User Guide
9
App DisplayDescription
Calibration Expired
Appears when a calibration has not been performed
in 24 hours. The system returns to the Initialisation
Phase.
No glucose reading can be displayed until
calibration is performed.
Battery Empty*
Appears once when your smart transmitter battery is
empty and needs to be charged.
No glucose reading can be displayed until the smart
transmitter is charged.
Actions
In the Initialisation Phase, you must perform
4 fingerstick calibration tests spaced 2 - 12 hours
apart. Display of glucose readings will resume after
the 2nd successful fingerstick calibration test.
Charge the smart transmitter immediately. Remove
the smart transmitter from your body before
connecting it to the power supply.