91
Eversense XL CGM User Guide
9
Alert Descriptions and Actions (continued)
Alerts
App DisplayDescription
Sensor Suspend
Appears when the system’s internal checks detect a
need to restart the Initialisation Phase for additional
calibrations. Glucose values will be displayed a few
minutes after the second successful calibration
during the Initialisation Phase.
Vibration Motor
Appears every 20 minutes when the vibration motor on
your smart transmitter can no longer provide on-body
vibe alerts. You will continue to get glucose readings up
to 72 hours after receiving the alert message. After
72 hours, you will receive a Transmitter Error Alert every
20 minutes until you replace the smart transmitter.
Actions
Re-initialisation of the system begins in 6hours. Contact Customer Support to have your smart
transmitter replaced immediately.