3.1 Specifications subject to restrictions depending on the combination
3.1.1 Restriction condition
1. Graphic cards
Depending on the type of graphic card used, the highest frame rate for live images may not be ensured.
2. Controller
The controller's installation software in the factory settings is in the scope of warranty. Malfunctions due to
changes in PC environment settings (BIOS changes), OS updates, and installation of third-party software by
the customer are not covered by the warranty. In addition, malfunctions after OS installation with the
recovery disk included in the controller are not covered by the warranty either.
3. Adaptors used in combination with the microscope
This product can be used with C-mount camera adapters marked with "V" or " " in the table below.
There may be some restrictions depending on the components of the microscope to be combined.
Camera adapter
U-TV0.5XC-3 U-TV0.63XC MVX-TV0.63XC-2 GX-TV0.7XC U-TV1XC
U-TV1XC-2
+U-CMAD3-2
MVX-TV1XC
- - - V V
V: Can be combined (thee may be some restrictions such as fundamental constraints)
-: Not recommended because the four corners of the image will be darkened due to mismatch with the
specifications of the product.
: Insufficient peripheral brightness may occur in the field of view depending on the components of the
microscope to be combined. It is recommended to use the shading correction and the partial cropping
function (ROI) of the image together.
4. A list of restrictions
Category Restrictions
1.
Fundamental
constraints
1. When the intermediate lens tube is long (see the example below), there may be
insufficient peripheral light in the field of view depending on the type of lens tube
and objective lens used together.
An example of a long intermediate lens tube:
l
Two or more intermediate lens tubes are combined and used together
l
BX3 series: a vertical illuminating light projection tube is used in
combination with the intermediate lens tube
l
IX3 series: IX73P2F or IX83P2ZF is used as a microscope body
2. When the AS (aperture stop) is narrowed, spot flare may become noticeable. This
flare becomes less noticeable when the AS (Aperture Stop) is opened.
3. When used in conjunction with a non-NIR compatible microscope, insufficient light
intensity or flare may be noticeable even with the IR cut filter removed.
2. Flare 1. If there is a high-brightness area in the screen that saturates the gradation, flare
may occur in the low-brightness area. It tends to be noticeable when the AS
(aperture stop) is narrowed to close to the minimum, and it will be less noticeable
when the AS (aperture stop) is opened. In addition, it will be less noticeable if
exposure amount is suppressed to the extent that the gradation is not saturated.