Page 28 © Boule Medical AB, March 2018. Article no. 1504496
3. Operation (Sample Analysis)
Analyzing Sample (Micro Pipette Adapter, MPA)
Note: Do not remove MPA device during sample aspiration or analysis. Removal prior to
completion of analysis may cause erroneous results.
Read section 4 on Venous Blood Sample Collection” before commencing.
X Analyze Venous Sample (Micro Pipette Adapter, MPA)
1 Enter Sample Information Follow instructions 15 under Analyzing Sample (Open
Tub e) to enter sample and ID information.
2 Preparing MPA device Pull out the MPA device. (The analyzer will give an
instruction to put back the loaded MPA device to start the
analysis cycle).
Remove the previous sample micropipette. (If applicable)
Place the adapter on the table.
3 Sample Collection
Once again, see section 4, “Venous Blood Sample Collection”
for this step.
4 Fill micropipette with venous
sample
Use the micropipette holder to grasp a micropipette (holding
it on one end and not the middle will facilitate lling of
blood).
Using your other hand tilt the sample vial so the blood nears
the opening of the tube.
Place the micropipette end into the sample vial and aspirate
blood via capillary action.
When the micropipette is completely lled, remove it from
the vial.
Wipe o any excess blood on the outside surface without
removing any blood from the inside of the capillary tube.
5 Micropipette insertion to
device and analyzer
Insert the micropipette into the MPA device as shown
above, using the micropipette holder.
Insert the MPA device into the analyzer which automatically
starts the analyzing sequence.
6 Sample Measurement The analyzer now switches to the sample analysis screen.
Sample ID1/ID2 and prole can be changed up until results
are displayed.
If any changes are made, press to save, and then
Conrm. Results will not be shown until change is
conrmed.
7 Results Displayed Sample results will be displayed.
Note: Do not remove MPA device during sample aspiration or analysis. Removal prior to
completion of analysis may cause erroneous results.