© Boule Medical AB, March 2018. Article no. 1504496 Page 25
3. Operation (Sample Analysis)
Analyzing Sample (Open Tube)
3 The aspiration time is approximately 10 seconds. After ~ 10 seconds the analyzer will time out
due to no detection of blood, and continue its cycle.
4 The background count should not be higher than the values shown in gure 28:
Rerun sample if values are not acceptable.
Analyzing Sample (Open Tube)
The following steps will guide the operator through analyzing a blood sample using the “Open Tube”
mode, which aspirates the blood sample through the sample probe.
X Sample Analysis Flowchart
Go to Start Menu.
Choose Sample and
Prole Type.
Aspirate patient sample by
pressing Start Plate.
Analyzer performs sample
measurements and clean cycle.
− Results are displayed.
− Sample analysis complete.
Enter in Sample ID1, Sample
ID2, and/or Operator ID.
X Analyzing Sample (Open Tube)
Figure 29: Start Menu
Figure 30: Sample Entry Keyboard
1 Enter Sample Analysis Mode Go to Start Menu.
2 Choose Sample type Choose Blood tab, in upper right-hand corner, for sample type.
3 Choose Prole type The analyzer can hold ≥ 20 dierent proles.
Choose prole by selecting the circle next to desired prole
type.
To see more proles use left and right arrows to scroll to
more prole types.