10. Analyzer Care and Maintenance
Cleaning
SECTION 10. ANALYZER CARE AND
MAINTENANCE
Cleaning
Daily Cleaning
The majority of the Exigo H400 systems cleaning procedures are automated to keep routine
cleaning to an absolute minimum, increase the longevity of the analyzer and decreases maintenance
procedures.
Clean the sample probe using a paper tissue moistened with a 70% alcohol solution.
Remove possible traces of salt crystals or blood at the top of the sample probe and probe rinse
cup using a paper tissue moistened with the alcohol solution.
When necessary, gently clean the display and/or outside of the analyzer with a soft cloth,
slightly moistened with water and a mild soap. Dry carefully.
Automatic Cleaning
The Exigo system has been designed to clean internal components on a daily basis. The system uses
the enzymatic cleaner to ush and clean all components that come into contact with blood when in
standby or power-o mode. The analyzer remains lled with the cleaner until it is powered back
on or taken out of standby. This automatic daily cleaning increase the longevity of the analyzer and
decreases maintenance procedures.
Important: Follow your lab established biohazard barrier protection. This may include gloves,
lab coat and/or eye protection.
Orice Cleaning
Check with your local Service Contact for more information.
Monthly Cleaning
To secure the correct function of the instrument a Clot Prevention procedure taking approximately
15 minutes should be performed on a monthly basis.