8. Technology
Measurement of Eosinophils
Measurement of Eosinophils
4 part WBC dierential impedance based analyzer
The method is included the conventional 3 part dierential analysis together with an added EOS
measurement with a total analyzing time of about 3 minutes.
After the 3 part dierential analysis the instrument continues with an additional WBC dilution
but instead of lyse, EOS reagent is used to achieve the nal 1:400 dilution.
After an incubation time all cells are lysed except the eosinophils.
The EOS population is displayed as a separate size-distribution curve, see g below.
A settable discriminator is used to separate the EOS population from debris and to count the total
number and percentage in relation to the total WBC count.
Figure 119: EOS Dierential