10. Analyzer Care and Maintenance
Transport (Short-term and Long-term)
Figure 120: Maintenance Menu Figure 121: Cleaning Menu
Clot Removal
Check with your local Service Contact for more information.
Annual Cleaning
Boule Cleaning Kit Procedure
To increase the life of the analyzer’s internal tubing, the following cleaning procedure is strongly
recommended. The Boule Cleaning Kit procedure takes approximately one hour and 15 minutes to
complete.
X Cleaning Procedure
1 Select Main Menu, then Maintenance, and arrow over to next page to enter the Cleaning Menu.
2 Follow the instruction for the Boule Cleaning Kit to clean the analyzer. (Instructions for use are
supplied with the Boule Cleaning Kit solutions).
The Boule Cleaning Kit contains the following items:
Hypochlorite (2%)
Enzymatic cleaner
Detergent cleaner
Transport (Short-term and Long-term)
Relocation of analyzer (within the laboratory)
This section describes the procedure performed to move the analyzer over very short distances (From
table to table).
X Analyzer Relocation
1 Before Relocation
If the analyzer is in Standby mode do not unplug analyzer. Use the Power Down button in
the Maintenance Menu to turn the instrument o.